VETERANS AFFAIRS, DEPARTMENT OF › NETWORK CONTRACT OFFICE 23 (36C263)
H266--Sources Sought - Equipment Testing and Program Monitoring Services of Radiology and Quality Programs - VA Black Hills Health Care System
Key facts
- Response deadline
- Dec 3, 2026
- 12:00 PM UTC-06:00
- Set-aside
- Not stated
- Notice type
- Sources sought
- Posted on SAM.gov
- Nov 25, 2026
- Archives Dec 5, 2026
Description
Show the full descriptionShow less
Page 1 of 38 Page 14 of 39 Page 1 of 38 The VA Black Hills Health Care System Fort Meade VA Medical Center, Rapid City CBOC, and Hot Springs VA Medical Center locations is conducting market research in accordance with RFO 17.204-1 Exercise of Options for Equipment Testing and Program Monitoring Services of the Radiology and Quality Programs as described in the Preliminary Statement of Work below. This is a SOURCES SOUGHT NOTICE only looking for businesses to provide the services listed in the Preliminary Statement of Work below for market research purposes. No proposals are being requested or accepted with this notice. THIS IS NOT A SOLICITATION FOR PROPOSALS OR FINAL PRICING AND NO CONTRACT SHALL BE AWARDED FROM THIS NOTICE. This notice shall not be construed as a solicitation or as an obligation on the part of the Department of Veterans Affairs. Responses to this notice will be treated only as information for the Government to consider as part of their market research efforts. The information provided will be used by the Government in developing its acquisition strategy regarding possible set aside for Service-Disabled Veteran-Owned, Veteran-Owned, and other socio-economic categories of small business. Interested parties are responsible for adequately marking proprietary, restricted or competition sensitive information contained in their response.End of the preview. Use “Show the full description” above to read all of it.
Page 1 of 38
Page 14 of 39
Page 1 of 38
The VA Black Hills Health Care System Fort Meade VA Medical Center, Rapid City CBOC, and Hot Springs VA Medical Center locations is conducting market research in accordance with RFO 17.204-1 Exercise of Options for Equipment Testing and Program Monitoring Services of the Radiology and Quality Programs as described in the Preliminary Statement of Work below.
This is a SOURCES SOUGHT NOTICE only looking for businesses to provide the services listed in the Preliminary Statement of Work below for market research purposes. No proposals are being requested or accepted with this notice. THIS IS NOT A SOLICITATION FOR PROPOSALS OR FINAL PRICING AND NO CONTRACT SHALL BE AWARDED FROM THIS NOTICE. This notice shall not be construed as a solicitation or as an obligation on the part of the Department of Veterans Affairs.
Responses to this notice will be treated only as information for the Government to consider as part of their market research efforts. The information provided will be used by the Government in developing its acquisition strategy regarding possible set aside for Service-Disabled Veteran-Owned, Veteran-Owned, and other socio-economic categories of small business. Interested parties are responsible for adequately marking proprietary, restricted or competition sensitive information contained in their response. The Government does not intend to pay for the information submitted in response to this notice and parties responding will not be entitled to payment for direct or indirect costs incurred in responding to this notice.
The North American Industry Classification System (NAICS) code for this requirement is 811210, Electronic and Precision Equipment Repair and Maintenance, with an SBA Small Business Size Standard of $34.0 Million.
If you are a vendor capable of providing the requested information and required services described below with competitive pricing, send your information with a description of proof of capability and answers to all questions to: [email protected] on or before Thursday, December 3, 2026 at 12:00 PM Central Time.
Please provide answers as appropriate to the following questions in the table below with your response to this Sources Sought. Failure to respond accurately to the following questions may affect the acquisition strategy. Failure to respond to the questions at all, or not respond to select individual questions, will result in your response being determined as non-responsive.
1.Â
Please provide all socio-economic categories of your firm (e.g., SDVOSB, VOSB, WOSB, Large Business, 8a, etc.) as well as your firm s SAM Unique Entity ID number (replaces DUNS number) and Government Contract POC.
2.Â
State whether the requested services may be ordered against a government contract awarded to your organization (e.g Federal Supply Schedule (FSS), General Services Administration (GSA), etc.).
3.Â
State if subcontracting is contemplated for this requirement and what percentage of the work will be subcontracted and for what tasks.
4.
Provide estimated lead time to deliver a fully functioning solution as described in the Preliminary Statement of Work below.
5.
Complete the ITEM INFORMATION directly below this section for market research pricing for the services described in the Preliminary Statement of Work.
ITEM INFORMATION
ITEM NUMBER
DESCRIPTION OF SUPPLIES/SERVICES
QUANTITY
UNIT
UNIT PRICE
AMOUNT
0001
1.00
JB
$___________
$___________
Perform certification via X-ray equipment testing, dose rate testing, and performance monitoring on radiographic fluoroscopy X-ray equipment for the Black Hills Health Care System during the month of April 2027 in accordance with the Statement of Work.
Contract Period: Base
POP Begin: 12-15-2026
POP End: 12-14-2027
PRINCIPAL NAICS CODE: 811210 - Electronic and Precision Equipment Repair and Maintenance
PRODUCT/SERVICE CODE: H266 - Equipment and Materials Testing - Instruments and Laboratory Equipment
0002
1.00
JB
$___________
$___________
Perform leak testing of sealed sources for the Black Hills Health Care System during the month of April 2027 in accordance with the Statement of Work.
Contract Period: Base
POP Begin: 12-15-2026
POP End: 12-14-2027
PRINCIPAL NAICS CODE: 811210 - Electronic and Precision Equipment Repair and Maintenance
PRODUCT/SERVICE CODE: H266 - Equipment and Materials Testing - Instruments and Laboratory Equipment
0003
1.00
JB
$___________
$___________
Perform leak testing of sealed sources for the Black Hills Health Care System during the month of October 2027 in accordance with the Statement of Work.
Contract Period: Base
POP Begin: 12-15-2026
POP End: 12-14-2027
PRINCIPAL NAICS CODE: 811210 - Electronic and Precision Equipment Repair and Maintenance
PRODUCT/SERVICE CODE: H266 - Equipment and Materials Testing - Instruments and Laboratory Equipment
0004
5.00
JB
$___________
$___________
Perform certification via X-ray equipment testing, dose rate testing, and performance monitoring as requested on all newly acquired radiographic fluoroscopy X-ray equipment for the Black Hills Health Care System throughout the period of performance and also perform follow-up testing on existing equipment on an as-needed basis when repairs are made or equipment is changed out requiring re-testing in accordance with the Statement of Work.
Contract Period: Base
POP Begin: 12-15-2026
POP End: 12-14-2027
PRINCIPAL NAICS CODE: 811210 - Electronic and Precision Equipment Repair and Maintenance
PRODUCT/SERVICE CODE: H266 - Equipment and Materials Testing - Instruments and Laboratory Equipment
0005
7.00
EA
$___________
$___________
Travel Expenses to perform services in accordance with the Statement of Work. Not to exceed $2,500 per instance.
Contract Period: Base
POP Begin: 12-15-2026
POP End: 12-14-2027
PRINCIPAL NAICS CODE: 811210 - Electronic and Precision Equipment Repair and Maintenance
PRODUCT/SERVICE CODE: H266 - Equipment and Materials Testing - Instruments and Laboratory Equipment
PRELIMINARY STATEMENT OF WORK
Introduction
The Black Hills Health Care System, Facilities Management Department requires a Board Certified diagnostic radiological physicist to perform equipment testing and program monitoring services of the radiology and quality programs at the Fort Meade VA Medical Center, Rapid City CBOC, and Hot Springs VA Medical Center locations. The Board Certified diagnostic radiological physicist shall have the capability to support and service ALL listed radiographic/fluoroscopic x-ray equipment and also provide certification as requested of all newly acquired radiographic/fluoroscopic x-ray equipment throughout the period of performance of the contract and all exercised option years.
Requirements
General:
The Contractor shall provide a Board Certified diagnostic radiological physicist to perform equipment testing as specified in Paragraph 2.b. for all necessary equipment in April of each year (see Paragraph 2.d. Equipment Listing) and program monitoring services of the radiology and quality programs at the Fort Meade VA Medical Center, Rapid City CBOC, and Hot Springs VA Medical Center locations. Services requested also include leak testing of sealed radiation sources as specified in Paragraph 2.c. in April and October of each year for specified equipment (see Paragraph 2.d. Equipment Listing). Additional services include follow-up testing and certification on existing equipment to be done on an as-needed basis as we make repairs or change out equipment that requires certification prior to use.
Radiographic Fluoroscopic X-ray Equipment Testing, Dose Rate Testing and Performance Monitoring (All April frequency cycle items in list below)
Testing and monitoring will conform to current VHA National Health Physics Program (NHPP) guidelines as noted in the references below.
Testing will include but not be limited to:
Measurement of radiation output parameters, including beam filtration
Testing of automatic control systems such as automatic exposure controls of radiographic systems and automatic exposure rate controls of fluoroscopy systems
Assessment of image quality
Measurement of appropriate indices of patient doses or dose rates at clinical technique factors with comparison to national standards
Measurement of minimum detectable activity of the Thyroid Uptake unit;
Monitoring will include but not be limited to:
A review of the quality control monitoring performed by technologists. The design will conform to the ACR Technical Standard for Diagnostic Medical Physics Performance Monitoring of Radiologic and Fluoroscopic Equipment and the ACR Technical Standard for Diagnostic Medical Physics Performance Monitoring of Computed Tomography (CT) Equipment , available from the American College of Radiology.
Appropriate indices of doses to patients, as estimated by the diagnostic radiological physicist for the technique factors used at the facility, must be posted at or near the control console for each x-ray imaging device.
For radiographic equipment, the entrance skin exposure, air kerma, or dose for a patient of typical size must be posted for common radiographic procedures.
For fluoroscopic equipment, the entrance skin exposure rate, air kerma rate, or dose rate must be posted for a typical range of patient sizes for common modes of operation, including various image magnification factors, boost mode, if available, continuous fluoroscopy, and for pulsed fluoroscopy at the available pulse rates.
For fluoroscopic equipment, the dose per image for each mode of image recording shall also be posted for a patient of typical thickness.
Leak Testing of Sealed Sources (All April/October Frequency Cycle Items in Paragraph 2.d. Equipment Listing):
Testing and recordkeeping requirements. Each licensee who uses a sealed source shall have the source tested for leakage periodically. The licensee shall keep a record of leak test results in units of micro curies and retain the record for inspection.
Method of testing. The wipe of a sealed source must be performed using a leak test kit. The wipe sample must be taken from the nearest accessible point to the sealed source where contamination might accumulate. The wipe sample must be analyzed for radioactive contamination.
Test frequency. (1) Each sealed source (except an energy compensation source (ECS)) must be tested at intervals listed (April/October). The Radiation Physicists will provide the wipe test kits, the analysis and the required report.
Equipment Listing:
Page 1 of 38
Page 15 of 39
Page 1 of 38
Freq
Fac
Room #
Nomenclature
Manufacturer
Model
Serial Number
VA AEMS MERS ID
Apr
FM
217A-145
RADIOG-FLUORO IMAGING SYST (R/F)
GE Healthcare
OEC ELITE
FBXXTE02427
568 MX3387487
Apr
FM
CS201I-113
R/F UNITS: MOBILE, OEC ELITE CFD (CART)
GE Healthcare
OEC ELITE CFD
FBXXTE03024
568 MX3580117
Apr
FM
S222-113
R/F UNITS: MOBILE, OEC ELITE CFD (CART)
GE Healthcare
OEC ELITE CFD
FBXXTE03025
568 MX3580118
Apr
FM
CS201I-113
R/F UNITS: MOBILE, OEC ELITE CFD (C-ARM)
GE Healthcare
OEC ELITE CFD
X0034828007971
568 MX3580488
Apr
FM
S222-113
R/F UNITS: MOBILE, OEC ELITE CFD (C-ARM)
GE Healthcare
OEC ELITE CFD
X003482007868
568 MX3580489
Apr
FM
104-265
MRI
Siemens
Magnetom Sola
183215
MX2583351
Apr
FM
148-113
CT
Siemens
Definition AS
96286
MX665120
Apr
FM
154-113
Spec-CT
GE Healthcare
Optima NM/CT640
640Y25559
MX2037514
Apr
FM
143-113
Radiographic Systems
Samsung Electronics
GC85A
52RNM3RM900001J
MX2040652
Apr
FM
143-113
Radiographic Systems
Samsung Electronics
GC85A
55EVM3EM7000001W
MX2287241
Apr
FM
143-113
R/F System
Siemens
Axiom Luminous agile
61059
EE43067
Apr
FM
143-113
x-ray mobile
Samsung Electronics
GM85
55GFM31M400011K
MX1973107
Apr
FM
143-113
x-ray mobile
Samsung Electronics
GM85
55GFM31M400012E
MX1973112
Apr
FM
313OP2-145
xray Dental Intraoral
Sirona
Heliodent plus
17647
EE38836
Apr
FM
313OP3-145
xray Dental Intraoral
Sirona
Heliodent plus
31019665
MX663076
Apr
FM
317-145
xray handheld intraoral
Aseptico Inc
Nomad pro 2
1001091
MX1882089
Apr
FM
318-145
xray Panolypse
Planmeca
Promax S3 PAN
TPX356791
EE38863
Apr
FM
155-113
Dose Calibrator
Capintec Instruments
CRC-25R
250014
EE28935
Apr
FM
156-113
Thyroid Uptake System
Biodex Medical Systems
HIT-W151
KSA1590781
MX419236
Apr/Oct
FM
155-113
Cs137 .0001 curies
Capintec Instruments
CR-15E
S8117001-06
none
Apr/Oct
FM
155-113
Ba133 .000231 curies
Capintec Instruments
CR-178E
S8107001-04
none
Apr/Oct
FM
155-113
Co57 .99561 curies
RadQual
BM06E
BM06057E11278103
none
Apr
FM
127-113
Densitometers: Bone: X-Ray
GE Healthcare
Lunar IDXA
213130MA
MX3495174
Apr
RCBOC
Mobile
Mobile MRI
Siemens
Magnetom AERA
142349
MX2067566
Apr
RCBOC
1910
Radiographic System Digital
Phillips Corp
Digital Diagnost C90
22860092
MX2991885
Apr
HS
D020-12
Mobile Scanning Systems: CT
Siemens
SOMATOM DEFINITION AS
95402
EE43219
Apr
HS
B006-12
R/F System
Siemens
Axiom Luminous agile
61237
MX423229
Apr
HS
B002-12
Radiology System
Samsung Electronics
GC85A
52RNM3FM500001N
MX2047221
Apr
HS
B002-12
xray portable
Samsung Electronics
GM85
55GFM31M4000014D
MX1980227
Apr
HS
D112A-12
xray Panolypse
Planmeca
Promax S3 PAN
TPXV3062268
MX1655374
Apr
HS
D112B-12
xray Dental Intraoral
dentsply Sirona
Heliodent plus
63217
MX1958839
Apr
HS
D112C-12
xray Dental Intraoral
dentsply Sirona
Heliodent plus
63234
MX1958841
Apr
HS
D112E-12
xray Dental Intraoral
dentsply Sirona
Heliodent plus
63209
MX1958843
Apr/Oct
FM
155--113
CSs137 .000199 Curies
New England Nuclear
NES 356
3560180A-50
None
Apr/Oct
FM
155-113
Co60 .00005 curies
New England Nuclear
NES 354
3540579A-49
none
Apr/Oct
FM
155-113
Ba133 .00029 curies
New England Nuclear
NES 358
3580479A-02
None
Apr/Oct
FM
155-113
Co57
RadQual
BM06E-57
BM6057E12083117
None
Apr/Oct
FM
155-113
Co57
RadQual
BM01-10
BM011012088101
None
Training Requirements:
Appropriate vendor staff (as determined by the VA) performing the duties of the Statement of Work shall complete (print, review, and sign) the attached TMS 20939 No Access document. The Contractor shall send signed TMS 20939 documents to the Contracting Officer s Representative Jonathan Maruszewski at [email protected].
Security:
The Contractor shall adhere to only the following IT security sections contained in the IT Contract Security Clause below (VHA Handbook 6500.6):
GENERAL
Working Hours:
Anticipated hours of work will be Monday through Friday from 7:00 am 4:00 pm. Additional hours must be coordinated through the COR/POC. Work will not be performed on weekends or federal holidays.
Observed Federal Holidays:
New Year s Day Martin Luther King Birthday
Washington s Birthday Memorial Day
Juneteenth Independence Day
Labor Day Columbus Day
Veteran s Day Thanksgiving Day
Christmas Day
References:
NHPP X-ray imaging guidance : http://nhpp.med.va.gov/NHPP_Diagnostic_Radiology_Info.asp
ARC Technical Standard for Diagnostic Medical Physics Performance Monitoring of Computed Tomography (CT) Equipment.
ARC Technical Standard for Diagnostic Medical Physics Performance Monitoring of Radiological and Fluoroscopic Equipment.
NCRP Report #102 Medical X-ray, Electron Beam and Gamma-Ray Protection for Energies up to 50MeV (Equipment Design, Performance and Use)
NCRP Report # 147 Structural Shielding Design for Medical X-ray Imaging Facilities
NCRP Report # 99 Quality Assurance for Diagnostic Imaging Equipment
NRC NUREG 1556, Volume 9 Program-Specific Guidance About Medical Use Licenses.
Joint Commission Compliance Checklist: Diagnostic Imaging Standards: EC.02.01.01 - PI.02.01.01
Contacts
ALISHA MILANDER · primary
701-239-3700 x3176
Attachments
BidLookup records the links this notice gives. It has not downloaded or checked the files.
Record dates
- Posted on SAM.gov: Nov 25, 2026
- The agency’s publication date, as SAM.gov states it.
- First recorded by BidLookup: Nov 25, 2026
- The UTC day of the first read of SAM.gov’s listings that stored this notice.
- Last confirmed: Nov 25, 2026
- The UTC day of the latest read that still listed this notice. It does not mean the notice is unchanged or still open on SAM.gov now, and it is not the index’s last-read time.
Published by the agency on SAM.gov as notice d71be9354dc54d59a85ffe60d07b1008. What these dates mean
Record
- Solicitation number
- 36C26327Q0137
- Search the index for this number
- NAICS
- 811210 · Electronic and Precision Equipment Repair and Maintenance
- Product/service code
- H266 · EQUIPMENT AND MATERIALS TESTING- INSTRUMENTS AND LABORATORY EQUIPMENT
- Place of performance
- FORT MEADE, SD 57741
- Contracting office
- Saint Paul, MN 55101
- SAM.gov notice ID
- d71be9354dc54d59a85ffe60d07b1008