Gas Chromatograph with a Thermal Conductivity Detector and Flame Ioniztion Detector with Methanizer
Key facts
Response deadline
Sep 21, 2026
12:00 PM UTC-04:00
Set-aside
SBASmall business (total)
Small Business Set Aside - Total
Notice type
Solicitation
Posted on SAM.gov
Sep 17, 2026
Archives Oct 6, 2026
Description
***Amendment 0001**** Amendment 0001 provides resposnes to questions received in response to the solicitation. Q.1. I am reviewing Notice ID 1333ND26QNB640601 and would appreciate some clarification regarding the analytical requirements. Would you be able to share the analytes of interest, the required detection limits, and the expected concentration ranges for the system? A.1. This analyzer will be used for the analysis of binary gas mixtures in nitrogen or air balance gases. The analyses performed will include a wide range of gaseous standard reference materials including: 1 ppm methane in air and 250 ppm carbon monoxide in nitrogen, both analyzed using the flame ionization detector with methanation, and 0.5 % carbon dioxide in nitrogen analyzed using the thermal conductivity detector. This analyzer will also be used for analyses of ambient level CO in air mixtures, ranging from (40 – 200) ppb. Q.2. What types of samples will be tested? A.2. This analyzer will be used for analyses of high pressure gas mixtures in nitrogen or air balance gases. Q.3. What specific substances or compounds need to be detected or analyzed in these samples? If available, please also share any specific testing standards or requirements. With this information, we can better understand your application and recommend a suitable instrument configuration and quotation accordingly. A.3. As stated above, this analyzer will be used for the analysis of high pressure gas mixtures delivered at or below 25 psig.
Show moreShow less
The analyses performed will include a wide range of gaseous standard reference materials including: 1 ppm methane in air and 250 ppm carbon monoxide in nitrogen, both analyzed using the flame ionization detector with methanation, and 0.5 % carbon dioxide in nitrogen analyzed using the thermal conductivity detector. For improved precision, temperature ramping and stable detector response are required to achieve baseline resolution between components in air balance gas. In all cases the repeatability should be at or below 0.2 % relative area counts. Q.4. Regarding your software requirements, you cite Visual Basic Version 6.0. Would that software be able to generate an XML file to communicate with our software (CompassCDS)? A.4. The visual basic software was developed to support communication between the GC software and a 12-port dead-end VICI sampling valve. The software can retrieve data from a txt file generated by the GC software and output the data to excel. There is not an extension to the solicitation as a result of this amendment. The question period for this soliciation will close today, September 17,2026, at 4:00 PM ET. End of Amendment 0001 ******** See attached.
From SAM.gov’s daily data file, as published on Sep 29, 2026 (UTC). SAM.gov’s latest daily file no longer lists this notice.
The agency’s publication date, as SAM.gov states it.
First recorded by BidLookup: Sep 20, 2026
The UTC day of the first read of SAM.gov’s listings that stored this notice.
Last confirmed: Sep 20, 2026
The UTC day of the latest read that still listed this notice. It does not mean the notice is unchanged or still open on SAM.gov now, and it is not the index’s last-read time.
Description as published: Sep 29, 2026 (UTC)
The date of the daily data file this text came from.
Published by the agency on SAM.gov as notice c3d1218425364835b7b429d402a51245. What these dates mean